Agilent Technologies 8133A 3 GHz Pulse Generator Technical Specifications
Setting Standards
Agilent Technologies 8133A 3 GHz Pulse Generator Technical Specifications
The need for pulse generation is fundamental to any device characterization task. The ability to emulate the pulse conditions to which the device will be subjected is essential. This emulation should include both typical and worst case conditions. Such accurate emulation requires superlative signal integrity and timing performance.
Setting Standards
The Agilent 8133A 3 GHz Pulse Generator provides pulses with programmable period from 333 ps to 30 ns, full 3 GHz pulse capability on all channels. The pulse width can be programmed too, along with a delay or the interchannel delay. At these frequency ranges the transition time performance becomes critical less than 100 ps is specified, less than 60 ps is typical, so excellent signal integrity is assured. And last but not least, a typical jitter of 1 ps creates precise and accurate timing conditions.
Pulse and Data Functionality
Human Interface
The Agilent 8133A is available in four different configurations. The differences are timing capabilities and number of channels with different functionality.
8133A Channel 1 Channel 2 Width or Delay with Square
Option 001 Delay also in Width Mode
Option 002 Delay also in Width Mode 32-bit data PRBS 223-1 Divided Square
Option 003 Delay also in Width Mode Second Pulse Channel with Width or Delay with Square
Note: option 002 and 003 contain option 001
Specifications
Timing Characteristics
Delay Pulse Channel 1 Option 001
Pulse Channel 2 Option 003
Pulse / Data Channel 2 Option 002
Accuracy Width range 1 Resolution Accuracy Duty Cycle Resolution
Pulse mode: no
Add. Variable delay Channel 1 to Channel 2 Delay 4 Resolution Accuracy Phase
Square mode: 0.000 ns 10.000 ns 1 ps
Square mode: 0.000 ns to 10.000 ns Pulse mode: -5.000 ns to + 5.000 ns Square mode: -5.000 ns to 15.000 ns
° ° 0° to 3600° ° 0.1°, best case 1 ps
Resolution Skew8
Notes:
Repeatability: is typ. four times better
than accuracy.
External Clock
The external input signal determines the timebase of the instrument. The external period or frequency is measured and displayed, allowing the correct setting of Duty Cycle and Phase also in External Clock mode.
External Frequency Counter Period Range: 333 ps to 30.303 ns (300 ps to
Internal Clock Generation
The internal clock can be set in frequency or period mode
ZERO-DELAY REFERENCE
TIMEBASE External Input TRIGGER Output PULSE Channel 1 Output OPTION 001 PULSE Channel 1 Output (Option 001)
EXT Divide mode (divide by 1) Divide mode (divide by 1) PULSE mode SQUARE mode PULSE mode SQUARE mode
18 ns typical Variable Delay 0 to 10 ns Variable Delay -5 to 5 ns Variable Delay -5 to 15 ns
OPTION 002 (includes Option 001) PULSE / DATA Channel 2 SQUARE mode (divide by 1) Output (Option 002) PULSE / DATA Channel 2 DATA RZ mode Output (Option 002) PULSE / DATA Channel 2 DATA NRZ mode Output (Option 002) Output BIT 0 mode TRIGGER Output (Option 002 only) OPTION 003 (includes Option 001) PULSE Channel 2 PULSE mode Output SQUARE
2 ns typical Variable Delay 0 to 10 ns
Interchannel Timing Diagram
Frequency Range: 33.0 MHz to 3.0000 GHz (2
PULSE CHANNEL 1
Delay Delay has no period limitations. Fixed delay between TRIGGER Output to CHANNEL 1 Outputs: 18.8 ns nominal for trigger output divide by 1
Divide
External Input Divide
PULSE / DATA CHANNEL 2 (# 002)
Delay
Fixed delay from TRIGGER Output to CHANNEL 2 Outputs. Trigger on pulse: 18.8ns nominal for trigger output divide by 1
32 BIT
32 bit of programmable data are selectable.
PRBS: 223 -1, CCITT 0.151Norm
Trigger on Bit 0: - 2.0 ns before Bit 0
nominal
External Input Range: 33 MHz - 3 GHz Interface: ac-coupled Impedance: 50 W nominal Minimum Swing
Selects Data mode. Further selection between 32bit programmable data or PRBS is required.
PULSE CHANNEL 2 (# 003)
Divide
In Pulse and Square modes, the frequency of Channel 2 can be divided by 2, 4, 8, 16, 32 or 64. The minimum Channel 2 frequency is 1 MHz.
Level / Pulse Performance Characteristics
The output voltage can be set in amplitude / offset or high- / low level mode. Levels double when driving into open circuits.
Limit
Maximum high and low levels into 50 W can be limited to protect the device under test. Pushing the limit key declares present levels as limits which then cannot be exceeded as long as this mode is active.
Complement (not #002 PULSE / DATA CHANNEL 2) Normal / complement selectable Data (only #002 PULSE / DATA
CHANNEL2) Provides logical complement of data
Disable
Relays connect / disconnect outputs
Connector Type
SMA (f) for all outputs and inputs
TRIGGER Output
Bit 0 (PULSE / DATA CHANNEL 2 only)
GP-IB Capabilities
All modes and parameters are programmable. Operates according to IEEE standard 488.1 and 488.2, 1987. Conforms to the Standard Commands for Programmable Instruments (SCPI) 1992.0.
Trigger Channel Output
Rearpanel Connectors
Prop. delay External Input to Trigger Channel Output: 6 ns nominal for
Weight
Net: 21.5 kg (48 lb) Shipping: 29.0 kg (65 lb)
Disable
A relay connects / disconnects the trigger output
Dimensions
(HxWxD) 145 mm x 426 mm x 525 mm (5.7 in x 16.75 in x 20.65 in)
Divide
The trigger output frequency can be divided. Divide by: (1), 2, 4, 8, 16, 32, 64
Start / Stop Input
Minimum output frequency: 33 MHz (3
MHz typical)
Additional Features
Non-volatile Memory Current settings are saved on powerdown. Additionally 20 complete settings can be stored.
Ordering Information
Agilent 8133A 3 GHz Pulse Generator
Agilent 8120-4948 50 W Cable, SMA
(m-m)
Option 001 Delay CHANNEL 1 Option 002 PULSE / DATA CHANNEL 2
Agilent 8710-1582 Torque Wrench 5
Agilent 8493A Series Attenuators Option 003 PULSE CHANNEL 2 Agilent 11667B Power Splitter Option 004 Standard Instrument Agilent 15435A Transition Time Note: Option 002 and Option 003 contain Option 001 Converter 150 ps
Agilent 15432B Transition Time Accessories Agilent 1250-1462 Adapter SMA (m)
to SMA (f) Converter 250 ps
Agilent 15433B Transition Time
Converter 500 ps
Agilent 15434B Transition Time
Converter 1000 ps
Agilent 15438A Transition Time
Converter 2000 ps
Related Agilent Literature
Agilent Family of Pulse / Pattern Generators, brochure, p / n 5980-0489E
By internet, phone, or fax, get assistance with all your test & measurement needs Online assistance: www.agilent.com / find / assist
Phone or Fax United States: (tel) 1 800 452 4844 Canada: (tel) 1 877 894 4414 (fax) (905) 206 4120 Europe: (tel) (31 20) 547 2000 Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840 Latin America: (tel) (305) 267 4245 (fax) (305) 267 4286 Australia: (tel) 1 800 629 485 (fax) (61 3) 9272 0749 New Zealand: (tel) 0 800 738 378 (fax) 64 4 495 8950 Asia Pacific: (tel) (852) 3197 7777 (fax) (852) 2506 9284 Product specifications and descriptions in this document subject to change without notice.
|