NDIR TYPE INFRARED ANALYZER
This NDIR analyzer features high accuracy, multiple functions easy operation through microprocessor. also utilizes mass flow detector noted high sensitivity reliability. This analyzer used only measuring environmental pollution also used various processes and/or experiments. This product, combined with sampling system (ZSP), passed type-approval test based Weight Measure Act.
FEATURES
microprocessor provides high accuracy, multiple functions easy operation. maximum components calibrated with built-in automatic calibrator (option). Using signal from zirconiz sensor (ZFK3) from other sensor, analyzer able output value converted into (option). Alarm function emit high/low limit contact output also available (option). Measuring range selected using external single (option). Zero span calibrated with high accuracy, simply pressing calibration keys. self-diagnosis function included. This analyzer utilizes mass flow detector featuring high sensitivity reliability. standard measuring ranges with range ratio 1:20. Three different components measured simultaneously connecting zirconia sensor (ZFK3), addition one-and two-component type sensors. Besides standard measurement type, sample switching type differential flow type also available.
SPECIFICATIONS
General items
Power supply: 100V 50/60Hz 115, 220V also available request
Note: Refer "Code symbols"
Power consumption: 125VA max. Ambient temperature: +45°C Ambient humidity: less Enclosure: Steel casing, indoor-use
Dimensions Mass{weight}: Approx. 24kg Finish color: Munsell 2.5Y 8.4/1.2 Indication: 4-digit concentration 4-digit sub-indication Output hold: Output hold prior manual/auto calibration possible. Hold ON/OFF selectable. Standard requirements sample gas: Temperature 50°C (dehumidification saturation less) Dust 0.3µm less Pressure 9.8kPa{0.1kgf/cm2} less (flow rate 0.5R/min) Standard control sample gas: Calibration Interfere control saturation Warm-up time: About hours (after power About hours sample switching type Material gas-contacting parts: Sample cell; stainless steel, neoprene rubber Infrared-ray transmitting window; CaF2 sapphire Internal tubing; Teflon tube, silicone tube, toaron tube inlet/outlet, purge inlet size: internal thread) NPT1/4 internal thread
EDS3-110c
Date Feb. 2001
Purge flow rate: 0.5R/min Purging required when measuring contained atmosphere range less than 50ppm. other cases, purging should made necessary. Scope delivery: Analyzer, test report, power fuse, cloth cleaning infrared-ray transmitting window Mounting method: Flush mounting (vertical mounting) Installation condition: Install analyzer place exposed direct sunlight radiation from high temperature object. Avoid vibration, select clean place free from corrosive and/or combustible gases. installing outdoors, provide suitable casing cover protect analyzer from wind, rain, etc. Measuring system: Non-dispersion infrared-ray absorption method, single light source, double beam, sample switching system Measurable component: (carbon monoxide) Measuring range:
range range, range, range (*4) [ppm] [ppm]
Select from Max. range rate within 1:10 range<2nd range<3rd range<4th range
Standard type
Measuring system: Non-dispersion infrared-ray absorption method, single light source, double beam Measurable components measuring range: Standard single-component analyzer
Measurable component Carbon monoxide Carbon dioxide Nitrogen monoxide Sulfur dioxide Methane Min. measuring range [ppm] Measuring range
20mA linear, step output which changes every seconds Repeatability: Within full scale (*3) (within full scale) Linearity: Within full scale Zero drift: Within 0.5% full scale/week (*3) (within full scale/week) Span drift: Within 1.5% full scale/week (*3) (within 2.5% full scale/week) Response time: Within seconds (for response) (*3) (within seconds) Measured flow rate: 0.1R/min (reference gas) 0.1R/min (sample gas) Sample switching relay drive output: ON/OFF signal second interval (for solid state relay drive)
Note: *(3) *(4) 2ppm range Sample switching type with 4-ranges available (Max. range ratio within 1:10). 4-range analyzer, optional functions remote range, range identification, alarm external hold available.
Output signal:
Page
Standard two-component analyzer
Measurable component component Nitrogen monoxide sulfur dioxide Carbon monoxide carbon dioxide Nitrogen monoxide carbon monoxide Min. measuring range [ppm] 100/0 100/0 250/0 Refer "Code symbols" Measuring range
Differential flow type
(Gas analyzer suited measurement modes, absolute concentration concentration difference) Measuring system: Non-dispersion infrared-ray absorption method, single light source, double beam, differential flow system Measurable components measuring range:
range [ppm] Carbon dioxide Carbon monoxide range [ppm]
Measuring range:Refer table. Output signal: 20mA (allowable load resistance, less), linear Repeatability: within 0.5% full scale (*1) (within full scale) Linearity: Within full scale Zero drift: Within full scale/week (*1) (within full scale/day) (*2) (within full scale/week) Span drift: Within full scale/week (*1) (within full scale/day) (*2) (within full scale/week) Response time: Max. seconds (for response) including substitution time; time differs with length sample cell. Measured flow rate: 0.25R/min
Note: *(2)
-100 +100 1000 None, 100, 200, 250, None, 200, 250, None, None,
Note: There restrictions reference conditions.
Output signal:
range 20mA range 20mA
Note: Linear output
range less than 50ppm 50ppm range
Sample switching type
(Sample switching type analyzer suited measurement concentrations eliminating effects interfering components.)
Repeatability: Linearity: Zero drift: Span drift: Response time:
Within 0.5% full scale within full scale Within full scale/week Within full scale/week Max. seconds (for response) including substitution time Measured flow rate: 0.25R/min (reference gas) 0.25R/min (sample gas)
Optional specifications
(There added request. Refer "Code symbols".) correction output: private sensor used correcting measured concentration into value standard concentration. obtaining exhaust standard value, concentration residual oxygen concentration exhaust measured simultaneously, then corrected according following equation. (Application this equation mandatory exhaust standard.)
where Concentration after correction measured concentration measured concentration standard concentration Block diagram (Example measurement)
signal signal measured signal according private sensor (ZFK) external meter DC/0 O2). corrected [ppm]
output signal: 20mA linear correction output signal: 20mA linear Output provided each components Alarm output: Upper limit alarm Contact output, SPDT (1c) contact Contact capacity, 250V (resistive load) Lower limit alarm Contact output, SPDT (1c) contact Contact capacity, 250V (resistive load) Remote range changeover: Range changeable external signal. Range changeover input signal: (minimum range selection input) Range identification signal output: Contact output, SPST (1a) contact Contact capacity, 250V (resistive load) External hold: Output hold possible with external input signal. External hold signal input: (hold input) Average value output: Moving average value average value outputted intervals 1-hour 4hour (only 1-output type 4-hour average value available). Output signal: 20mA linear.
Automatic calibration: Zero span automatically calibrated present cycle. Calibrating supplied operating external electromagnetic valve. Calibration channels: components calibrated simultaneously. Zero calibration point: Fixed (air point span point zirconia meter) Span calibration point: 100% full scale Calibration start: Built-in timer remote start signal Output hold during calibration: Possible Calibration flow mode: Zero Zero span Zero span span Zero span span (O2) Zero span span span (O2) Calibration flow time: Settable from seconds Calibration cycle: hours (1-hour units) days (1-day units) Calibration failure alarm: Provided when fault occurs during auto calibration. Contact output: Under calibration; SPST (1a) contact, contact capacity 250V (resistive load) Calibration failure; SPST (1a) contact, contact capacity 250V (resistive load) Electromagnetic valve drive; SPST (1a) contact, contact capacity 250V (resistive load) Remote start: Remote start signal Voltage input (start 100msec pulse input)
FUNDAMENTAL DIAGRAM
(Standard type)
Trimmer Reference cell Interference compensating detector
Distribution cell (interference filter)
Infrared light source (single light source) Motor Rotary sector (one-point chopper) Sample inlet Sector synchronous signal meter (Option) Sample voltage amplifier Sample outlet Gain changeover
Contact input, output Alarm output
Indicator Display Operating Output, 20mA
Range identification signal Contact output Various during calibration correction average Remote range input value value External hold input (Option) (Option)
(Differential flow type)
Distribution cell (interference filter) Reference Trimmer Reference outlet Reference cell inlet Interference compensating detector
Infrared light source (single light source) Motor Rotary sector (one-point chopper) Sample inlet Sector synchronous signal Sample voltage amplifier Sample outlet Gain changeover
Contact input,output Alarm output Range identification signal Calibration output drive signal Remote range input External hold input (Option)
Indicator Indication Operating Output, 20mA
FUNDAMENTAL DIAGRAM
(Sample switching type)
Trimmer Interference compensating detector Reference outlet
Distribution cell (interference filter)
Reference cell Reference inlet
Infrared light source (single light source) Motor Sample cell Rotary sector (one-point chopper) Sample inlet Sector synchronous signal voltage amplifier Sample outlet Gain changeover
Contact input,output Alarm output Range identification signal Calibration output drive signal Remote range input External hold input (Option)
Indicator Relay drive signal Indication Operating Output, 20mA
Description sample switching system
measured divided into two, flow paths, high performance converter provided convert carbon monoxide into carbonic acid gas. Here, carbon monoxide measured eliminated into cell high-sensitivity infrared analyzer. other flow path, directly into other cell analyzer. These flow paths changed over changeover valves every seconds means changeover valve drive signal transmitted from analyzer. carrying this changeover cyclically, output analyzer varies with concentration carbon monoxide measured gas. obtained variation width measured value enables improving ratio eliminating effect interfering components plus zero drift.
Temperature controller CO/CO2 converter
filter
ZRG3 Reference cell
measured filter Pump
Flowrator
Sample switching type analyzer Sample cell
Reference
Relay Output signal
CODE SYMBOLS
(Standard single-component analyzer)
Description Measuring method Single-component analyzer Single-component analyzer (approved Weight Measure Act)
Measurable components Other non-standard items component, range 5ppm 10ppm 15ppm 20ppm 25ppm 50ppm 100ppm 200ppm 250ppm 500ppm 1000ppm 2000ppm 5000ppm 100% Other non-standard items Power supply 100V 115V 220V 220V Output signal 20mA 50/60Hz 50/60Hz 50/60Hz 50/60Hz linear linear Range identification Note: Details average value alarm, remote range, output depend 16th external hold 17th digits. Average value output includes moving average value output.
Note: When selecting range, sure refer table "Measuring range". specifications Dioxin analyzer 200ppm max. different some points.
Piping connection
Optional functions Automatic Average value calibration output
Optional functions indication indication Range identification Note: case indication, (external (ZFK connection) alarm, remote range corrected outputted analyzer standard specification. connection) Standard type analyzer ranges 10/25%)
Note: Corrected designated 18th digit. External analyzer required linear output.
(Standard single-component analyzer) (cont'd)
component, range range without 10ppm 15ppm 20ppm 25ppm 50ppm 100ppm 200ppm 250ppm 500ppm 1000ppm 2000ppm 5000ppm Other non-standard items Average value output time Without 1-hour moving average value output 4-hour moving average value output 1-hour average value output 4-hour average value output
Description Note: range>1st range Range rate: within 1:20
Note: (10) Average value output (option) designated only 12th digit.
Average value output object component component Instantaneous value corrected
Note: (11) Average value output (option) designated only 12th digit. (12) selection 4-hour moving average value output 4-hour average value output, only 1-output type available.
Note: (13) designation "Z", data corrected standard value must data code. Without Other non-standard items (specify within 19%) Measuring kind Atmospheric Combustible exhaust (Dioxin included) Converter exhaust Other non-standard items Non-standard spec. Other non-standard items Quick response Note: (14) designation "Z", components specifications should attached.
Note: (15) Data should data code. (16) Quick response type available only when density more.
(Standard two-component analyzer)
Description Measuring method Two-component analyzer Two-component analyzer (approved Weight Measure Act)
Measurable components (1st component) (2nd component) Other non-standard items component, range Code Note: Code Range value Code Range value Range value 2000ppm 5000ppm Other non-standard items 1000 range selection, sure refer table "Measuring range". min. range: 250ppm
component, range Code Note: Range value Code Range value Code Range value 2000ppm 5000ppm 1000 Other non-standard items range selection, sure refer table "Measuring range". min. range: 250ppm 50/60Hz 50/60Hz 50/60Hz 50/60Hz Piping connection
Power supply 100V 115V 220V 220V
Output signal linear 20mA linear Other non-standard items Optional functions Automatic Average value calibration output Range identification Note: Details average value alarm, remote range, output depend 16th external hold 17th digits. Average value output includes moving average value output.
Note: This function applies components. Optional functions indication indication Range identification Note: case indication, (external (ZFK connection) alarm, remote range corrected outputted analyzer standard specification. connection) Standard type analyzer ranges 10/25%)
Note: (10) Corrected designated 18th digit. (11) External analyzer required linear output. component, range Code Note: Code Range value Code Range value Range value 2000ppm Without 5000ppm Other non-standard items 1000 (12) range>1st range range rate: within 1:20
(Standard two-component analyzer) (cont'd)
component, range Code Code Range value Without 1000 Note: (13) range>1st range
Description Code Range value 2000ppm 5000ppm range rate: within 1:20 Range value Other non-standard items
Average value output time Without 1-hour moving average value output 4-hour moving average value output 1-hour average value output 4-hour average value output
Note: (14) Average value output (option) designated only 12th digit.
Average value output object component component component Instantaneous value corrected Instantaneous value corrected
Note: (15) Average value output (option) designated only 12th digit. (16) selection 4-hour moving average value output 4-hour average value output, only 1-output type available.
correct standard value Note: (17) designation "Z", data Without must data code. Other non-standard items (specify within 19%) Measuring kind Atmospheric Combustible exhaust (Dioxin included) Converter exhaust Other non-standard items Non-standard spec. Other non-standard items Quick response Note: (18) designation "Z", components specifications should attached.
Note: (19) Data should data code. (20) Quick response type available only when density more.
(Sample switching type)
Measuring method Sample switching system Measurable components range 2ppm 5ppm 10ppm 25ppm 50ppm 100ppm Other non-standard items
Description
Power supply 100V 115V 220V 220V Output signal 20mA
50/60Hz 50/60Hz 50/60Hz 50/60Hz linear linear
Piping connection
Optional functions Automatic Range identification calibration alarm, remote range, external hold
[Unit: ppm]
range/3rd range/4th range range range range Measuring kind Atmospheric Other non-standard items Non-standard spec. Other non-standard items
range 2ppm
range
range 10ppm
range 25ppm range 50ppm
Note: designation "Z", components specifications should attached. Note: Data should data code.
(Differential flow system)
Measuring method Differential flow system Measurable component (Carbon dioxide) (Carbon monoxide)
Description
range 50ppm 100ppm 50ppm 100ppm 200ppm 250ppm Piping connection Note: Voltage output signal only digits, (1st range: +1V) (2nd range:
Power supply 100V 115V 220V 220V
50/60Hz 50/60Hz 50/60Hz 50/60Hz
Output signal linear 20mA linear Optional functions Automatic Range identification, calibration alarm, remote range, external hold
range [Unit ppm] Without range 350±50 1000 range 350±100 range 50ppm range 100ppm range 200ppm range 250ppm Measuring kind Atmospheric Other non-standard items Non-standard spec. Other non-standard items Note: designation "Z", components specifications should attached. Note: Data should data code.
NDIR TYPE INFRARED ANALYZER
<Standard single-component analyzer measuring range>
Measuable component range measuring range
5ppm 10ppm 15ppm 20ppm 25ppm 50ppm 100ppm 200ppm 250ppm 500ppm 0.1% 0.2% 0.5% 100%
Remarks: Standard measuring range Consult regarding manufacture availability, price delivery period. manufactured
<Standard type two-component analyzer measuring range>
100ppm 100ppm 200ppm 250ppm 500ppm 0.1% 0.2% 0.5% 100%
200ppm 250ppm 500ppm Remarks Scope manufacture range
range type available provided range ratio component within 1:20. (Max. 2000ppm analyzer) 250ppm 500ppm 1000ppm 1000
Remarks range type available provided range ratio component within 1:20. (Max. 2000ppm analyzer)
Remarks (1): Scope manufacture range Scope manufacture range (maximum range) range type available provided range ratio component within 1:20.
OUTLINE DIAGRAM
(Unit:mm)
NOTE* mounted when reference used.
Reference inlet* Purge inlet
Reference outlet* Sample inlet Sample outlet
NPT1/4
Mounting size
panel
External terminal Approx.230 OPTION OPTION AUTO M3.5 screw
12hole Mounting screw should used.
Cable inlet (with rubber grommet)
Operation panel
from
61.5
Rock bracket
Front panel
Side panel
Power supply terminal (inside)
Cable inlet (with rubber grommet)
External connection diagram
20mA (instantaneous value) 20mA (moving average value) (option) 20mA (instantaneous value) 20mA (moving average value) (option) External hold
OPTION
sensor input 20mA (instantaneous value) component corrected value 20mA component corrected value 20mA FAULT component Remote range input (option)
component
component Remote range input (option)
component
component Range identification contact output (option)
component Range identification contact output (option) Remote range input (option)
Remote range input (option)
OPTION component Upper limit alarm contact output (option)
AUTO component Lower limit alarm contact output (option) Calibration contact output
Auto calibration failure contact output
Zero contact output component Lower limit alarm contact output (option)
Pump ON/OFF contact
component Upper limit alarm contact output (option)
Span gas-1 contact output
Remote start input
Upper limit alarm contact output (option)
Lower limit alarm contact output (option)
Span gas-2 contact output
Relay drive output signal (sample switching type only) Relay drive output signal (Differential sample switching type only)
Span gas-3 contact output
Note: relay drive output signal terminal, only sample switching type differential flow system used. Terminal block should used position AUTO-CAL terminal block.
Power supply (installed main unit)
Ground Power supply
Private zirconia sensor <option> This sensor used together with ZRG. Measuring method: Zirconia system Measurable component measuring range:
Measurable component Oxygen range range
Repeatability: Within 0.5% full scale Linearity: Within full scale Zero drift: Within full scale/week Span drift: Within full scale/week Response time: Approx. (for response) Measured flow rate: 0.25R/min
Note: Zirconia system, principle, produce measuring error relative concentration versus combustible concentration. Also, corrosive (SO2 250ppm more, etc.) affect life sensor.
Indication: Temperature indication (LED) Temperature alarm output: Contact output, SPST (1a) contact Contact capacity, 220V (resistive load) Dimensions 190mm Mass{weight}: Approx. Finish color: Munsell Y7/1
CODE SYMBOLS
Description Measuring method Zirconia system Zirconia system, (approval Weight Measure Act)
inlet/outlet size: Rc1/4 Power supply: 126V 240V 50/60Hz Enclosure: Steel casing, indoor application
Power supply 126V 50/60Hz 240V 50/60Hz
OUTLINE DIAGRAM
(Unit:mm)
OXYGEN ANALYZER
TEMP
POWER FUSE3A
7-M4
inlet
172.3
outlet
External connection diagram
screw terminal)
power supply, 50/60Hz
Output
Alarm output
Caution Safety *Before using this product, sure read instruction manual advance.
Head office
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan http://www.fesys.co.jp
Sales Div. International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187 http: //www.fic-net.co.jp
Information this catalog subject change without notice.
Printed Japan

Other recent searches


W78E54 - W78E54   W78E54 Datasheet
W78C54 - W78C54   W78C54 Datasheet
TPS3619 - TPS3619   TPS3619 Datasheet
MSP430 - MSP430   MSP430 Datasheet
LY251 - LY251   LY251 Datasheet
LTB93-B101P012BP - LTB93-B101P012BP   LTB93-B101P012BP Datasheet
LNJ816C83RA - LNJ816C83RA   LNJ816C83RA Datasheet
HAF1003 - HAF1003   HAF1003 Datasheet
DS1280 - DS1280   DS1280 Datasheet
DS05-20874-7E - DS05-20874-7E   DS05-20874-7E Datasheet
DB315C - DB315C   DB315C Datasheet
DB318C - DB318C   DB318C Datasheet
DB320C - DB320C   DB320C Datasheet

 

Privacy Policy | Disclaimer
© 2012 Datasheet Archive